Description
How Do I Connect My Solar & WiFi Smart Residual Chlorine Meter and Gateway to YIERYI APP ?
APP Connection
* Note : The solar water quality tester must be paired with a gateway in order to be used, our link is delivered with 1 gateway.
1.Connect the power, press and hold the “RESET” button for 5 seconds, the indicator light will be on and the red light will be flashing.
2.Open YIERYI APP, click Add Device.
3.Select the gateway control. Multi-function Gateway (Zigbee)
4.Input the correct Wi-fi password
5.Confirm the indicator light is blinking, select fast blinking.
6.Search for devices.
7.Click Finish.
8.Click Add Sub-device.
9.Press and hold for 5 seconds to blink the indicator light.
10.Confirmation light is blinking fast.
11.Searching for devices.
12. Connection is successful.
Yieryi WiFi Smart Residual Chlorine Meter Solar Floating CL PH EC TDS ORP SALT Swimming Pool Water Quality Monitor Tester with Gateway
【Solar & Floating CL PH EC TDS ORP SALT TEMP Water Quality Tester 】Solar & WiFi Smart Remote Control Swimming Pool Residual Chlorine Tester, which is a multi-functional water quality tester, scan the QR code to download the free Android or IOS app (APP: YIERYI Smart) and install it. Add Solar & WiFi Residual Chlorine Tester and WiFi hub on APP, connect Solar & WiFi Residual Chlorine Tester to 2.4GHz WiFi, then you can monitor the changes of water quality anytime, anywhere, and you can know the P Residual Chlorine/PH/EC/TDS/ORP/SALT/TEMP of the swimming pool water in a more accurate, quicker, and more stable way.
【Expanding Residual Chlorine Water Quality Monitoring System】–You can customise the device name and even have multi-account management and multi-device management. Expandable to multiple Solar & WiFi Smart Remote Control Swimming Pool Residual Chlorine Tester, build a smart monitoring system for your pool or multiple pools to take over the pool monitoring tasks. In addition, the product comes with an automatic temperature compensation function for better measurements, avoiding different temperatures affecting the measurement and ensuring accurate measurement data.
【Analyse Monitoring Data Anytime】–You can easily export the measurement data of Solar & WiFi Smart Remote Control Swimming Pool Residual Chlorine Meter, and view the historical data (monthly, weekly, daily) at any time, so that you can analyse the measurement data and save time. data, saving time and avoiding frequent recording and data loss.
【Scenario Setting / Alarm Mode for Exceeding Limit Values】–Solar & WiFi Smart Remote Control Swimming Pool Residual Chlorine Tester has an abnormal alarm mode, you can set the upper and lower limits of the measurement values (monthly, weekly, daily) at any time, so that you can analyse the measurement data and save time. You can set the upper and lower limits of the measurement values (only one value can be set at a time), when it reaches or falls below the set value, the remote monitor will send out an abnormal alarm in the APP to remind you to pay attention to the abnormal situation in time. You can even create smart scenarios, add tasks to send notification alerts, choose the alert method when the device status changes, and set the alarm value freely!
【Environmental Mode of Multi-Function Charging】–Solar water quality tester with 2 charging modes (solar/USB charging), which makes the product more convenient and long-lasting, and avoids replacing batteries frequently, making your life more environmentally friendly.
Frequently Asked Questions :
◆1.Why the residual chlorine value is 0 ?
The pool pH value should be within 6.5-8 and ORP value should be within 487-840, the residual chlorine value will be displayed normally.
If the ORP value is within 487-840, the residual chlorine value will be displayed normally, but if any one of these two values is out of the range, it will not be displayed normally.
◆2. Why can’t I connect to WiFi?
Make sure the device and your mobile phone are in the same WiFi network environment, make sure WiFi – 2.4GHZ.
◆3. Dear, if you have any questions about the product, please contact us, we have the video of removing the electrode cover before use, the video of instruction to connect the app, the calibration video of PH and ORP calibration (using calibration solution), even we have professional technical engineers to answer your questions.
BLE-YL01 Specification :
1.Measurement Range :
*PH : 0.0-14.0 pH
*CL : 0.0-4.0 mg/L
*ORP : -999-+999mV
*SALT : 0-999ppm; 1000-9990ppm
*EC : 0-2000μS/cm; 2000-9990μS/cm; 10.01-19.99mS/cm
*TDS:0-999ppm; 1000-9990ppm
*TEMP: 0.0-50.0℃; 32.0-122.0°F
2.Resution :
*PH : 0.1pH
*CL : 0.1mg/L
*ORP : 1mV
*SALT : 1ppm;10ppm
*EC : 1uS/cm;10uS/cm; 0.01ms/cm
*TDS : 1ppm;10ppm
*TEMP : 0.1℃/°F
3 . Precision :
*PH :±0.1pH
*CL:±0.1mg/L
*ORP:±15mV
*SALT:+2%F. S
*EC:±2%F. S
*TDS:±2%F. S
*TEMP:±0.1℃
4 . Product Features : Exception alerts;Data export;Remote monitoring;Solar/USB charging ;Multi-account management;Multi-device management;Name customization Setting of upper and lower value limits;Automatic temperature compensation function
Monthly, weekly and daily historical data view at any time
5 . Standby : The phone exits the APP and automatically enters standby mode,Open the APP, the device works normally for measurement.
6 . Product size : Length21cm*Width16.5cm
7 . Packing list : Mainframe*1 ; PH4.00 Calibration Solution*1 ; PH7.00 Calibration Solution*1 ; PH10.01 Calibration Solution*1 ; ORP Calibration Solution*1 ; Instruction Manual*1 ; Packaging Box*1
Set the upper limit value/lower limit value of the value :
* Note : If the test result is out of the setting range, the number will be displayed in red.
Data Export :
* Note : You can choose daily/weekly/monthly export Excel, export in the form of a table sent to the mailbox, send the mailbox and click download.
* Steps to export data:
1.Click Export
2.Input the mailbox
3.Open the mailbox to download
Scenario Setting / Overlimit Value Alarm :
* Note : Select message reminder, when the value is higher/lower than the alarm value APP will pop up the alarm message.
* Setup steps:
1.Create smart scene
4.Add task to send notification reminder
2.Select when the device status changes
5. Select the reminder method
3.Set the alarm value
6.Save
Calibration Method :
* Ph Calibration
1.Pour 7.00/4.00/10.01 calibration solution into the cup respectively.
2. Open the APP to enter the PH interface (the current device has been bound to operate).
3. Select PH standard (EU standard).
4. Put the electrode into the 7.00 solution, wait for the value to stabilise, click calibration, automatically identify the calibration solution
7.0 appears, complete 7.00 calibration.
5. Put the electrode into the 4.00 solution, wait for the value to stabilise, click calibration, the calibration solution will be recognised automatically.
4.0 appears, complete 4.00 calibration.
6. Put the electrode into the 10.01 solution, wait for the value to stabilise, click on calibration, the calibration solution is automatically recognised.
10.0 appears, complete 10.01 calibration.
* ORP Calibration
1. Pour 222mV calibration solution into the cup.
2. Put the electrode into the 222mV solution, enter the ORP calibration interface, wait for the value to stabilise, click calibration.
Adjust the value to 222, display calibration in progress, 222 appears, complete 222 calibration.
* EC Calibration
1. Pour 1413us/cm calibration solution into a cup, and put the electrode into the 1413 solution.
2. Enter EC calibration interface, wait for the value to be stable, click calibration, adjust the value to 1413, display calibration in progress.
1413 appears, complete 1413 calibration.
3. Pour the 12.88ms/cm calibration solution into the cup. Place the electrode into the solution and wait for the value to stabilise.
Click calibration, adjust the value to 12880, display calibration in progress, 12.88 appears, complete 12.88 calibration.
* Maintenance and care of the electrode
1. PH electrode should avoid long-term immersion in acidic fluoride solution. The electrode avoids contact with silicone oil.
2. When cleaning the electrode bulb, it should not be placed directly under the strong water current, but should be placed in pure water and rinsed gently.
3. When measuring, we should avoid cross contamination between solutions, so as not to affect the accuracy of measurement.
4. The electrode bulb is contaminated (pollutants: organic metal oxides, organic oils and fats, tree polymers).
Protein blood cell precipitates, pigment substances) will slow down the response speed, so according to the nature of the pollutants should be
According to the nature of the pollutant, the electrode should be cleaned with the appropriate solution to restore the performance of the electrode.
Reviews
There are no reviews yet.